How to Calibrate Dyes on Your Quant Studio 5: A Step-by-Step Guide

How to Calibrate Dyes on Your Quant Studio 5: A Step-by-Step Guide

Today, we're going to learn how to calibrate dyes on a machine called the Quant Studio 5. This machine is used in molecular biology, which is a fancy way of saying it helps scientists study tiny things like cells and genes. But don't worry, I'll break it down so it's easy to understand. Let's get started!

What You'll Need:

  1. A Quant Studio 5 machine
  2. Calibration plates (These are special plates with different colors, or "dyes," on them. For the Quant Studio 5, you'll need two types of plates.)

Step-by-Step Guide:

Step 1: Understand the Calibration Plates

There are two types of calibration plates you'll need for the Quant Studio 5. Each plate has a different mix of dyes on it. Here's what they look like:

Plate Dyes
Plate 1 FAM, VIC, ROX, SYBR, TAMRA
Plate 2 ABY, JUN, MUSTANG PURPLE, NED, CY5

Step 2: Start the Calibration Process

  1. On your Quant Studio 5 machine, go to "Settings."
  2. Then, click on "Maintenance and Service."
  3. Finally, click on "Calibrations."

Step 3: Choose the Right Calibration

There are different types of calibrations you can do:

  • ROI (regions of interest): This makes sure the machine is looking at the right parts of the plate.
  • Uniformity: This checks that the machine's camera is working properly.
  • Dye history: This keeps track of the different colors the machine has seen.

For this guide, we're focusing on dye calibration.

Step 4: Input the Plate Details

After selecting the plate you're using (either Plate 1 or Plate 2), you'll need to input the lot numbers and part numbers. These numbers are usually found on the box the plates came in.

Step 5: Start the Calibration

Now, you can start the calibration process. The machine will do this automatically. All you have to do is click "Start."

Step 6: Check Your Results

Once the calibration is done, you can check your results in the "History reminders" section. This will show you how well the calibration went.

Important Tips:

  • Be careful: If you make a mistake during calibration, you can't go back. The machine won't work until you get a new calibration plate. So, make sure you're ready before you start!
  • Prepare your plate: Before you start, your dye plate needs to be thawed for about 15 minutes. Then, you'll need to shake it for about five seconds and spin it down. This gets it ready for the machine.
  • Check your results: After the calibration, you'll see different colors (or "bands") on your results. If these bands are very tight, that means the calibration went well!

And that's it! Remember, calibrations usually last a couple of years, so you won't have to do this all the time. But now, if something happens to your machine, you'll know what to do.